Nutricia Scc3297 Flocare Infinity Pump Alarm manual
Unknown · 10 Q&As

Nutricia Scc3297 Flocare Infinity Pump Alarm

Ask AI

— answers from the official manual

Answers from the official manual.

Common questions

Common Questions

10 total
1

How often should maintenance be conducted on this device?

The manufacturer recommends an inspection at an authorized service center once every two years and replace the feeding sets every 24 hours to maintain accuracy (Page 10-13).

2

How do I activate the 'Low Battery' alarm feature?

This function is automatically enabled; when remaining battery shows one hour left, it will blink and start beeping every two seconds reminding of low charge condition (Page 14).

3

How do I switch off the pump safely?

Press and hold the ON/OFF key for 2 seconds. This will cause a continuous alarm sound and the pump will shut down, retaining all feeding program data and administered volume since the last clearance (Page 11).

4

What happens when I press the 'START/STOP' key?

When pressed once, it temporarily pauses the pump in hold mode until pressing again for normal operation or for three minutes before a two-tone alarm sounds prompting you to start it ('PUSH STRT' displayed on screen). (Page 11)

5

How do I clear the volume delivered by the pump?

Press and hold the 'CLR' key until the screen shows '0 ml', setting total volume delivered to zero. This operation resets memorized data, putting the flow rate at default values for new programming sessions (Page 13).

6

How do you prime the pump set correctly?

Use the FILL SET function by pressing and holding this button to automatically fill a new or empty feeding tube while deactivating alarms to avoid accidental triggering. Release button once initiated until full (Page 12-13).

Show 4 more questions

Full Manual

19 pages
Page 1

PTIT-CHFR-BE-CH-LUESDE-AT-CH-LUBE-NL UK-IE

![image 1](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile1.png)

![image 2](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile2.png)

![image 3](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile3.png)

![image 4](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile4.png)

![image 5](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile5.png)

![image 6](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile6.png)

![image 7](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile7.png)

![image 8](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile8.png)

![image 9](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile9.png)

![image 10](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile10.png)

![image 11](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile11.png)

![image 12](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile12.png)

![image 13](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile13.png)

![image 14](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile14.png)

![image 15](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile15.png)

![image 16](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile16.png)

![image 17](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile17.png)

![image 18](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile18.png)

![image 19](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile19.png)

![image 20](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile20.png)

![image 21](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile21.png)

![image 22](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile22.png)

![image 23](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile23.png)

![image 24](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile24.png)

![image 25](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile25.png)

Page 2

![image 26](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile26.png)

![image 27](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile27.png)

![image 28](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile28.png)

![image 29](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile29.png)

![image 30](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile30.png)

![image 31](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile31.png)

![image 32](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile32.png)

![image 33](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile33.png)

![image 34](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile34.png)

![image 35](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile35.png)

![image 36](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile36.png)

![image 37](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile37.png)

![image 38](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile38.png)

![image 39](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile39.png)

![image 40](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile40.png)

![image 41](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile41.png)

![image 42](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile42.png)

![image 43](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile43.png)

![image 44](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile44.png)

![image 45](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile45.png)

![image 46](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile46.png)

![image 47](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile47.png)

![image 48](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile48.png)

![image 49](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile49.png)

![image 50](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile50.png)

![image 51](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile51.png)

Page 3

![image 52](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile52.png)

![image 53](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile53.png)

![image 54](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile54.png)

![image 55](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile55.png)

![image 56](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile56.png)

![image 57](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile57.png)

![image 58](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile58.png)

![image 59](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile59.png)

![image 60](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile60.png)

INSTRUCTIONS FOR USE

For enteral use onl Read this manual before using the pump

![image 61](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile61.png)

![image 62](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile62.png)

![image 63](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile63.png)

![image 64](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile64.png)

![image 65](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile65.png)

![image 66](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile66.png)

![image 67](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile67.png)

![image 68](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile68.png)

![image 69](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile69.png)

![image 70](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile70.png)

![image 71](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile71.png)

![image 72](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile72.png)

BE-NL • INHOUD Algemene productinformatie 5 Voorbereiding 5 Service en garantie 5 Gebruiksaanwizing 6 Alarmfuncties en veiligheidsvoorzieningen 10 Onderhoud 10 Problemen en oplossingen 11 Instelmodus 12 Technische specificaties 14

DE-AT-CH-LU • INHALT Allgemeine Informationen 15 Vorbereitung 16 Service und Gewährleistung 17 Bedienungsanweisung 17 Alarmfunktionen und Sicherheitstechnik 22 Instandhaltung 22 Fehlersuchtabelle 23 Sstemeinstellungen 24 Technische Daten 25 Bestellinformation 26

ES • ÍNDICE Introducción 27 Preparación 27 Servicio  garantía 27 Instrucciones de funcionamiento 28 Funciones de alarma  prestaciones de seguridad 32 Mantenimiento 32 Diagrama de solución de problemas 33 Modo de configuración 34 Especificaciones técnicas 36

Page 4

FR-BE-CH-LU • CONTENT

Introduction 37 Précautions d’emploi 37 Service après-vente et garantie 37 Instructions d’utilisation 38 Alarmes et sécurités 42

Entretien 42 Résolutions des problèmes 43 Mode de paramétrage 44 Spécifications techniques 45

IT-CH • CONTENUTI

Introduzione 47 Preparazione 47 Servizio e garanzia 47 Istruzioni per l’uso 48 Allarmi e funzioni di sicurezza 52 Manutenzione 52 Risoluzione dei problemi 53 Funzione di parametrizzazione 54 Specifiche tecniche 55

PT • ÍNDICE Introdução 57 Preparação 57 Serviço e garantia 57 Instruções de funcionamento 58 Funções do alarme e características de segurança 62 Manutenção 62 Quadro para soluções e problemas 63 Modo de configuração - Set up 64 Especificações técnicas 65

UK-IE • CONTENT

Introduction 67 Preparation 67 Service and Warrant 67 Operating instructions 68 Alarm functions and safet features 72 Maintenance 72 Problem solver chart 73 Set up mode 74 Technical specifications 75

Appendix A: Guidance and manufacturer’s declaration Electromagnetic compatibilit 78

![image 73](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile73.png)

  • a
  • Figure 1

    d

    f c

  • g
  • h
  • e

  • b
  • Figure 3

    | | |---| | |

    Figure 4

    Figure 2

    ![image 74](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile74.png)

    ![image 75](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile75.png)

    Page 5

    i j

    | | | |---|---| | | |

    ![image 76](Nutricia_20FloCare_20Infinity_20-_20User_20manual_images/imageFile76.png)

    k

    Figure 5a

    Figure 5b

    |A
    B
    | |---|

    Figure 5c Figure 5d

    ZMNL400336_AW.indd 4 3/18/10 3:47:31 PM

    #### BE-NL POMPSPECIFICATIES:

  • Fig. 1:
  • a Pompdeur
  • b Scherm (LCD)
  • c Toetsenbord
  • d Rotor
  • e Stroomopwaartse druksensor
  • f Stroomafwaartse druksensor
  • g Vake voor pompset
  • h Luchtsensor
  • Fig. 2:

  • i Gebruiksaanwizing
  • j Bevestigingspunt voor paalklem
  • k Aansluiting voor adapter
  • Fig. 3: Paalklem
  • Fig. 4: Adapter
  • Fig. 5a-5d: De pompset vullen en

    inbrengen

    #### DE-AT-CH-LU KOMPONENTENLISTE:

    Page 6

  • Fig. 1:
  • a Pumpentüre
  • b Anzeigefeld (LCD)
  • c Tastatur
  • d Rotor
  • e Vorgeschalteter Drucksensor
  • f Nachgeschalteter Drucksensor
  • g Anschluss für das Pumpsegment
  • h Luftsensor
  • Fig. 2:

  • i Bedienungshinweise
  • j Anschluss für Schraubhalterung
  • k Buchse für den Netzanschluss
  • Fig. 3: Schraubhalterung
  • Fig. 4: AC/DC Adapter/Ladegerät
  • Fig. 5a-5d: Einführung des

    Pumpsegments

  • tema
  • h Sensor del aire
  • Fig. 2:
  • i Instrucciones de uso j Receptáculo para el fiador del

    polo

    k Toma para la conexión a la uni-

    dad cargadora de enchufe

  • Fig. 3: Fiador del polo
  • Fig. 4: Cargador
  • Fig. 5a-5d: Inserción del equipo de

    bomba

    #### FR-BE-CH-LU LISTE DE COMPOSANTS:

    Page 7

  • Fig. 1:
  • a Porte
  • b Ecran (LCD)
  • c Clavier
  • d Rotor
  • e Capteur de pression amont
  • f Capteur de pression aval
  • g Site insertion cassette
  • h Détecteur d’air
  • Fig. 2:

  • i Instructions d’utilisation
  • j Site insertion support
  • k Connexion de l’adaptateur
  • Fig. 3: Support
  • Fig. 4: Chargeur Fig. 5a-5d: Insertion de la tubulure
  • #### IT-CH ELENCO DEI COMPONENTI:

  • Fig. 1:
  • a Sportello pompa
  • b Visualizzatore a cristalli liquidi (LCD)
  • c Tastiera
  • d Rotore
  • e Sensore della pressione in ingresso
  • f Sensore della pressione in uscita
  • g Ricettacolo per l’inserto del set
  • h Sensore dell’aria
  • Fig. 2:

  • i Istruzioni per l’uso
  • j Ricettacolo per il sistema di fissaggio alla piantana
  • #### ES LISTA DE COMPONENTES: Fig. 1:

  • a Compuerta de la bomba
  • b Panel indicador (LCD)
  • c Teclado
  • d Rotor
  • e Sensor de presión ascendente
  • f Sensor de presión descendente
  • g Receptáculo para insertar el sis-
  • Page 8

    k Connettore per l’alimentatore

  • Fig. 3: Sistema di fissaggio alla piantana
  • Fig. 4: Caricabatterie Fig. 5a-5d: Inserimento del set
  • #### PT LISTA DE COMPONENTES:

  • Fig. 1:
  • a Porta
  • b Ecrã (LCD)
  • c Teclado
  • d Rotor
  • e Sensor de pressão ascendente
  • f Sensor de pressão descendente
  • g Receptor para colocação do sistema
  • h Sensor de ar
  • Fig. 2:

  • i Instruções de utilização
  • j Receptor para colocação do suporte
  • k Conexão para o carregador
  • Fig. 3: Suporte
  • Fig. 4: Carregador
  • Fig. 5a-5d: Colocação do sistema de

    alimentação na bomba

    #### UK-IE LIST OF COMPONENTS:

  • Fig. 1:
  • a Pump door
  • b Displa panel (LCD)
  • c Kepad
  • d Rotor
  • e Upstream pressure sensor
  • f Downstream pressure sensor
  • g Receptacle for pump insert
  • h Air sensor
  • Fig. 2:

  • i Instructions for use
  • Page 9

  • j Receptacle for pole clamp
  • k Socket for connection to the plug charger unit
  • Fig. 3: Pole clamp
  • Fig. 4: Charger
  • Fig. 5a-5d: Filling and Insertion of the

    pump set

    ORDERING INFORMATION Flocare

    #### pump Western Europe corporate code 35676 Flocare

    #### Infinit

    ®

    Infinit

    Charger

    ®

    Pole clamp Instructions for use

    Optional accessories Please refer to our local Nutricia contact for availabilit of Flocare

    accessories, as there are: feeding sets, carr bag, PDMS/data cable, nurse call, service suitcase and other.

    Infinit

    ®

    MANUFACTURER Nutricia Medical Devices b.v. Schiphol Boulevard 261 1118 BH Schiphol Airport The Netherlands

    #### INTRODUCTION

  • • The Flocare® Infinit™ pump is a small, lightweight pump for both portable and bedside use. The pump is based on the rotar peristaltic principle and is intended for enteral use onl.
  • PREPARATION

  • • Check the integrit of the pump. Do not use the pump if it is damaged. If a technical failure occurs or if the pump is dropped, the pump should be checked b a qualified technician.
  • • Donotusethepumpinareaswherethereisariskofexplosionse.g.inthepresenceofflammableanaesthetics.
  • Page 10

  • • If the pump has been stored for an period of time, it should be plugged into the mains to recharge the batter before commencing enteral feeding. The batter will be completel charged after approximatel 6 hours.
  • • The Flocare® Infinit™ pump should onl be used in combination with the appropriate Flocare® Infinit™ pump set! (consult our local Nutricia sales representative for information on availbale pump sets).
  • • This pump operates in an orientation, making it ideal for ambulator use.
  • • Check the position of the feeding tube, as advised b our healthcare professional, before commencing tube feeding.
  • • Pump fed patients should be regularl monitored and supervised. Specific patient groups require consistent and controlled administration of enteral nutrition as well as simultaneous application of medication (e.g. insulin administration). In these cases, regular and frequent checks, as determined b the attending healthcare professional, should be carried out to ensure correct administration of nutrition throughout the therap period. Using the Infinit™ pump’s DOSE function is recommended in these cases (see section “To set a Dose”).
  • • For bedside use, the multi-position pole clamp (figure 3) can be attached to the pump with the screw provided. The pump can be fixed in an position (rotatable in 360°).
  • #### SERVICE AND WARRANTy

    The manufacturer recommends an inspection of the pump at an authorised service centre ever 2 ears. Onl authorised personnel should perform service work on Infinit™ pumps. Please contact our local Sales Organisation / Nutricia Subsidiar for all service and repair of pumps (see address at the back of the booklet).

    Limitations of warrant Solel for the benefit of the original buer/user, Nutricia Medical Devices B.V, warrants all new Flocare® Infinit™ pumps, of its manufacture to be free from defects in material and workmanship, excluding normal wear and tear, and will replace or repair, at its service facilit or other location designated b Nutricia Medical Devices B.V, an Flocare® Infinit™ pump returned to it within thirt-six (36) months of original purchase b the buer/user. Such repair or replacement shall be free of charge.

    Nutricia Medical Devices B.V warrants to the original buer/user, all repaired or replaced pumps to be free from defects in material and workmanship and will replace or repair such products, at its service facilit or other location designated b Nutricia Medical Devices B.V. Such repair or replacement shall carr a warrant of ninet (90) das from the date of repair or replacement or the balance of the new pumps warrant as described above, whichever is greater.

    THIS WARRANTy APPLIES ONLy TO FLOCARE® INFINITy™ PUMPS MANUFACTURED By NUTRICIA MEDICAL DEVICES B.V AND IS THE ONLy WARRANTy GIVEN WITH RESPECT TO THE PUMPS. NO WARRANTIES IMPLIED IN LAW, INCLUDING, BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITy AND FITNESS FOR PARTICULAR PURPOSE, SHALL APPLy. NUTRICIA MEDICAL DEVICES B.V WILL BE LIABLE, IN ANy EVENT, ONLy FOR THE PURCHASE PRICE OF THE DEFECTIVE PRODUCT, BUT NOT FOR ANy CONSEQUENTIAL DAMAGES.

    This Warrant ma not be modified, amended or otherwise changed, except b a written document properl executed b a corporate officer of Nutricia Medical Devices B.V

    THE WARRANTy IS VOID IF THE FLOCARE® INFINITy™ PUMP IS SUBjECT TO ABUSE, ACCIDENT, ALTERATION, MODIFICATION, TAMPERING, MISUSE OR THE UNAUTHORIZED REPAIR OR SERVICE IN ANy WAy WITHOUT PRIOR AUTHORIZATION FROM NUTRICIA MEDICAL DEVICES B.V. IN ANy EVENT, NUTRICIA MEDICAL DEVICES B.V.’S LIABILITy SHALL NEVER EXCEED THE ORIGINAL PURCHASE PRICE OF THE PUMP AND SHALL NOT EXTEND TO ANy CONSEQUENTIAL LOSS OR DAMAGE.

    OPERATING INSTRUCTIONS INSERTION OF THE FLOCARE® INFINITy™ PUMP SET

  • • Connect the set to the feed container as instructed on the packaging of the Flocare
  • ®

    Infinit

    pump set. Remove the dust cap from the step connector.

  • • The Flocare
  • ®

    Infinit

    pump sets are equipped with an “automatic free-flow protection”, as a consequence the set has no roller clamp. Fill the pump set completel with feed b gentl pinching on the side of the cassette that is marked with a drop (point A on figure 5a) or fill the pump set with help of the pump (see paragraph: filling the pump set).

  • • Open the pump door b pressing up on the lower wall and rotating the door upwards at the same time (see figure 5b).
  • • Position the looped section of the silicone tubing around the rotor. Stretching lightl, (see figure 5c) seat the cassette into the pump (see figure 5d).
  • • Close the pump door.
  • ON/OFF

    #### SWITCHING “ON”

    Page 11

    For proper pump operation, make sure the pump door is closed prior to switching the pump on. Press the “ON/OFF” ke for 2 seconds. The pump beeps and carries out a short self test, showing the pump serial number in 8 digits. Verif that all displa segments and smbols are active as shown in figure 6 (see next page). The pump displas the total volume delivered since the memor was cleared and switches in the hold mode read for programming.

    #### SWITCHING “OFF”

    ON/OFF

    Press the “ON/OFF” ke and keep it pressed for 2 seconds. A continuous alarm will be heard and the pump switches off. The feeding program (installed parameters) and total volume administered since the last clearance will be retained in the pump’s memor. If the pump shuts down due to low batter voltage, the memor will be retained for 24 hours.

    #### “HOLD” MODE

    START/STOP

    To temporaril pause the pump, or switch into “hold” mode whilst operating, press the “START/STOP” ke once. Three beeps are heard and the run smbol disappears. The programmed flow rate (ml/h), volume (DOSE=VOL) and the administered volume (ml) are retained.

    The “hold” mode is used to temporaril stop the flow of feed:

  • • to change the feeding program (installed parameters),
  • • to change the feed container,
  • • to administer medication without switching the pump off,
  • • to silence an alarm and correct problems. After 3 minutes a two tone audible alarm sounds and the message “PUSH STRT” appears in the displa. Press “START/STOP” to stop the alarm and to extend the hold mode b a further 3 minutes or press “START/ STOP” twice (3 beeps will be heard) to resume programmed settings.
  • #### FILLING THE PUMP SET

    ON/OFF

    d

    FILL SET

    The FILL SET function offered b the Infinit pump is used to fill a (new) feeding set with nutrition (or water as the case ma be). When the FILL SET function runs all alarms are deactivated helping avoid alarms being inadvertentl triggered (e.g. air alarm). The FILL SET function should onl be used to fill an empt (air-filled) feeding set. Use of the FILL SET function at an other time ma cause the pump to incorrectl calibrate. When the FILL SET function is initiated the pump will calibrate to the giving set.

    When the pump is in the hold mode: Press and hold the “FILL SET” ke for 2 seconds to activate the “FILL SET” maneuver. Release the button as soon as the pump generates a beep and starts pumping at a flow rate of approx. 700ml/h. During this maneuver the message “FILL SET” appears in the displa. The pump will automaticall stop when the Flocare

    Pack Infinit

    ®

    pump set is completel filled with feed. However this “FILL SET” maneuver can at antime be stopped b pushing the “FILL SET” ke a second time. The pump will return to the hold mode when the “FILL SET” maneuver is complete or stopped.

    #### DISPLAy

    The pump has a liquid crstal displa (L.C.D.) with large alphanumeric characters, smaller smbols, words and a back light. The following information can be found in the displa:

    • Flow rate (ml/h), volume (ml), is displaed through the large characters. Words below describe what the number relates to (rate, dose or volume). The pump also displas messages, for example “end of dose” will appear when the pump has finished delivering a single feed dose.

  • •The arcs around the run smbol rotate when the pump is running.
  • Page 12

  • •The remaining batter capacit (see paragraph: batter operation).
  • •Thewallplugsmbolindicatesthatthechargerispluggedin.
  • Figure 6

    The back light of the displa turns off 10 seconds after the last ke is pressed. The back light will switch on for 10 seconds when the charger is connected to the mains.

    #### PROGRAMMING THE PUMP

  • • The Flocare
  • ®

    Infinit

    pump can be used for continuous or dose feeding.

  • • Insert the pump set in the pump, switch the pump on and fill the pump set when necessar (see paragraph: filling the pump set).
  • • The displa shows the total volume delivered since the pump was last cleared.
  • • If necessar clear the volume delivered b pressing the “CLR” ke.
  • • The last installed flow rate (ml/h) is now displaed.
  • • The pump is now in the “hold” mode and is read for programming. CONTINUOUS FEEDING
  • • Adust the flow rate (ml/h) if required b using the “+” or “-” ke. Hold either ke down to change rapidl.
  • • If another parameter is indicated in the LCD, it will be necessar to first press the “ml/h” ke followed b the “+” or “-” ke to set the flow rate.
  • • The flow rate ranges from 1 - 400 ml/h in 1ml increments.
  • • The flow rate slows down and stops shortl at 50 - 125 - 250 ml/h.
  • • Press the “DOSE=VOL” ke and make sure the value is put at 0 ml indicated b the word “CONT”. This means the pump will run in a continuous mode.
  • • Start the pump b pressing “START/STOP”.
  • • The arcs around the word “run” start rotating.
  • • The flow rate (ml/h) is displaed whilst functioning.
  • + 1 ml/h-

    ON/OFF d d dml/h400ml/h

    START/STOP

    #### TO SET A DOSE

    + 1 ml/h-

    4000 ml+ 1 ml-

    ON/OFF d d d d dml/h START/STOP400ml/h

    DOSE=VOL

    From the hold mode, with the last flow rate (ml/h) displaed in the LCD:

    Page 13

  • • Set the flow rate b pressing the “+” or “-” ke. If another parameter is displaed:
  • • Press first the “ml/h” ke followed b “+” or “-” ke to adust the flow rate. The flow rate can be set between 1 and 400 ml/h, with increments of 1ml.
  • • Press the “DOSE=VOL” ke and install the volume to be administered with the “+” and “-” ke. The range goes from 1 - 4000 ml, with steps from 1ml.
  • • Start the pump b pressing “START/STOP”.
  • During functioning the following parameters can be found in the displa:

  • • B pressing the “ml/h” ke the “flow rate” is displaed.
  • • B pressing the “INFO” ke the total volume since the memor was cleared will be visualized.

    INFO

    When the required volume or dose has been delivered, “END OF DOSE” will appear and the pump will beep (or mute, depending upon the pumps’ configuration (see paragraph: set up mode).

    #### TO CHANGE THE FEEDING PROGRAM DURING FUNCTIONING

  • • Press the “START/STOP” ke to pause the pump.
  • • Change the program b pressing the required ke (ml/h or DOSE=VOL) and adust using the “+” or “-” kes.
  • • Restart the pump b pressing the “START/STOP” ke again. CLEARING THE MEMORy
  • All parameters and values can be cleared one after the other in the following wa:

  • • Press “START/STOP” to pause the pump.
  • • Select the parameter that needs to be cleared.
  • • Press the “CLR” ke to clear the memor. The parameter or value returns to its default value: ml/h = 0 ml/h VOL = cont = No dose set. Pump will feed continuousl until feed container is empt or pump
  • is stopped

    INFO = 0 ml = Volume delivered is cleared In order to keep a clear overview of the dail amount of feed delivered, clear the total volume delivered as each dail feeding schedule is started, as follows:

  • • Switch the pump “ON”
  • • Immediatel after the self-test the pump displas total volume delivered.
  • • Press the “CLR” ke. The total volume delivered returns to “0 ml”.
  • • The pump displas the previous programmed flow rate (ml/h).
  • • The pump is in the “HOLD” mode read for programming.
  • • When another feeding program needs to be started, or the pump will be used for another patient, all settings (rate and dose) can be returned to default and the volume delivered can be set at zero b simpl pressing the “CLR” ke and keeping it pressed for 2 seconds.
  • BATTERy OPERATION

  • • The Flocare
  • ®

    Infinit

    Page 14

    pump is equipped with a Lithium-ion batter.

  • • The batter smbol in the displa is automaticall replaced b the plug smbol when the adapter is plugged in. The “fuel gauge” then indicates that the batter is charging b displaing the segments in an ascending low to high pattern starting with the leftmost segment. This pattern continuousl repeats while the pump is charging.
  • • To check the status of the batter, disconnect the charger from the pump and turn the pump on. The bars between E and F (E=Empt, F=Full) represent the “fuel gauge” of the batter. Each bar is approximatel 1/4 of a full batter charge. If 2 bars appear the batter is half full and remaining operating time is approximatel 12 hours at a flow rate of 125 ml/h.

  • • In case of power failure the pump automaticall switches to batter powered operation.
  • • When during functioning the last block is gone, the batter smbol will blink to indicate that there is approximatel 1 hour of charge left. The displa will flash “BATT” ever 3 seconds, alternating with the active displa and the pump will beep ever 2 seconds to remind the user of the low batter charge condition. Plug in the charger to continue to run and recharge the batter.
  • • As the batter level decreases the pump ma not be able to deliver tube feeds at rates greater than 300 ml/h, even though some batter capacit remains. The pump will alarm and indicate “BATT”. If this happens, either
  • START/STOP

    d

    CLR

    recharge the batter or select a lower rate to complete the feed ccle.

  • • In case of batter failure, caused b, for example, an excessive temperature condition (which ma resolve itself with time) or b a complete batter failure, the E, F and batter smbol will flash. Turn the pump “off” and contact our healthcare provider or refer the pump to service.
  • #### CONNECTION TO AN EXTERNAL ALARM SySTEM

    The power connector at the side of the pump offers the option of connecting the pump to an external alarm sstem or Patient Data Monitoring Sstem (PDMS).

  • • The Flocare
  • ®

    Infinit

    Nurse Call (corporate code 35752) enables connection to an external alarm sstem. This external alarm sstem ma be with open or closed contact. Follow the instructions for use of the Flocare

    ®

    Infinit

    Nurse Call to connect the pump.

  • • The Flocare
  • PDMS Cable (corporate code 35776) enables connection to an external PDMS sstem. Follow the instructions for use of the Flocare

    Infinit

    ®

    Page 15

    Infinit

    PDMS Cable to connect the pump.

    ®

    ALARM FUNCTIONS AND SAFETy FEATURES In case an of the problems listed on the problem solver chart occur, the pump delivers an audible and visual alarm and stops working. The back light of the LCD automaticall switches on. Exception to this is the low batter alarm “BATT”, in this situation the pump continues working. Action in the event of an alarm: Check the tpe of alarm displaed b the large displa characters.

  • • Press the “START/STOP” ke to stop the audible and visual alarm.
  • • Correct the cause of the alarm as described in the table.
  • • Start the pump again b pressing “START/STOP”. The problem solver chart on the next page gives a clear explanation of the alarms.
  • MAINTENANCE

    Cleaning:

  • • Alwas unplug the pump prior to cleaning to avoid electric shock hazard.
  • • On a regular basis thoroughl clean all surfaces of the pump (including the sensors and rotor) with warm soap water, a 5% bleach solution in water, or a multipurpose disinfectant cleaner.
  • • The Flocare
  • ®

    Infinit

    pump ma be rinsed b holding under a stream of warm, clean water. Do not submerge the pump!

  • • Alwas maintain the rollers on the rotor in a clean state to ensure the spin freel.
  • • The adaptor normall does not require cleaning. When desired, a dr or slightl damp cloth ma be used to clean the outside surface of the adaptor. Make sure the adaptor is disconnected from the wall outlet.
  • ###### Safet notes:

  • • The manufacturer recommends an inspection of the pump at an authorised service centre ever 2 ears.
  • • If an fault occurs during use, or if the pump is dropped, it should be checked b authorised technical personnel prior to use.
  • • Replace the Flocare
  • ®

    Infinit

    pump set ever 24 hours to maintain deliver accurac and prevent the growth of harmful bacteria. Dispose of Flocare

    ®

    Page 16

    Infinit

    disposable sets properl, as required b local law.

  • • Do not use the Infinit alarm sstems to trigger actions related to secondar (electrical) medical devices (e.g. a volumetric or sringe pump).
  • • Do not use pump functions (e.g. the fill set function) for an other purpose than described in this manual, as this ma cause the pump to incorrectl calibrate.
  • PROBLEM SOLVER CHART

    Alwas follow the instructions below in case a problem occurs. Using methods other than those described ma cause the pump to function incorrectl.

    |NO SET

    Condition|•The set is not fitted or wrongl fitted in the pump.
    •The pressure sensor area is dirt


    Cause|•Stop the alarm b pressing “START/STOP”
    •Insert the Flocare® Infinit™ cassette into the pump as indicated on the blister packaging and close the door.
    •Restart the pump.
    •Clean the sensors, reinsert the cassette in the pump and restart the pump.


    Correction| |---|---|---| |PUSH STRT|•The pump has been untouched in hold mode for 3 minutes or more.|•Stop the alarm and prolong the hold mode with another 3 minutes b pressing “START/STOP” ke.
    •Program the pump and start it b pressing the “START/ STOP”.
    | |END OF DOSE|•The pump administered the installed dose volume.|•Turn the pump off b pressing the “ON/OFF” ke and hold it during 2 seconds or
    •Clear the memor of the total volume administered (see paragraph: clearing the memor), reprogram a new feeding schedule and start the pump b pressing “START/ STOP”.
    | |PROG|•No flow rate is installed. Flow rate 0 ml/h|•Make sure the correct flow rate is programmed.| |DOOR|•The door is not correctl closed.|•Make sure the door is properl closed prior to starting a feeding program.| |AIR|•The air sensor detected an empt pump set.
    •The air sensor area is dirt.
    •The feeding set is not inserted correctl.
    |•Replace the empt feeding reservoir and continue feeding. If necessar prime the set.
    •Make sure the air sensor is clean.
    •Make sure the feeding set is properl inserted in the pump.
    | |BATT the pump remains working|•The batter capacit is too low. The pump is not able to deliver highl viscous fluids at high flow rates with the present charge level of the batter.|•Connect the adaptor to the pump and mains and charge the pump for approx. 6 hours. During charging the pump can be used.| |Batter, “E” and “F” flashing|•Batter failure.|•Turn the pump “off”, contact our healthcare provider or refer the pump to service.| |FILL SET|•The pump is priming the set.|•Press the “FILL SET” ke another time to stop the pump and bring it back in the hold mode.| |OCC IN|•The pump detected an upstream occlusion between the pump and the feeding bag.
    •The pressure sensor area is dirt.
    |•Stop the alarm b pressing the “START/STOP” ke.
    •Remove the feeding set out of the pump and check the permeabilit b flushing the line.
    •Re-insert the feeding set in the pump and restart.
    •Clean the sensors, reinsert the cassette in the pump and restart the pump.
    | |OCC OUT|•The pump detected a downstream occlusion between the pump and the patient.
    •The pressure sensor area is dirt.
    |•Stop the alarm b pressing the “START/STOP” ke.
    •Check the permeabilit of the feeding tube b aspirating liquid via the medication port.
    •Clean the sensors, reinsert the cassette in the pump and restart the pump.
    | |OCC OUT (repeated)|•Calibration not et completed on the current feeding set|•Stop the alarm b pressing the “START/STOP” ke.
    •Start the pump b pressing the “START/STOP” ke and let it run onl briefl.
    •Stop the pump b pressing the “START/STOP” ke insuring that there has been no occlusion out alarm.
    •Remove the feeding set from the pump and reinsert the feeding set in the pump.
    •Restart the pump b pressing the “START/STOP” ke.
    |

    |Condition LOCK

    |Cause

    •Onl the current feeding schedule is allowed for this patient. Another feeding program is not allowed b our healthcare professional.|Correction

    •The programming feature is blocked in the set up mode of the pump. Ask our healthcare professional to modif this setting.| |---|---|---| |ER01 - ER99|•The self test detected an electronic error|•Turn the pump “off”, make sure the pump door is closed and switch the pump back “on”. If the error persists, contact our healthcare provider or refer the pump to service.| |No plug smbol visible, while the pump is connected to the mains.|•The wall outlet doesn’t work.
    •The adaptor is damaged.
    |•Connect the pump to another wall outlet.
    •Contact our healthcare professional or Nutricia subsidiar to replace the adaptor.
    |

    SET UP MODE

    The set up mode is used • to set the alarm level,

  • • to lock the kepad,
  • • to switch the audible alarm off when a dose is administered,
  • • to switch the light of the LCD permanentl on when connected to the mains,
  • • to enable connection with a Nurse call or Patient Data Monitoring Sstem (PDMS).
  • This set up mode ma onl be entered b healthcare professionals and authorised personnel trained to use this application of the pump. The patient or his/her relatives without permission of the phsician, dietician, nurse or other licensed practitioner, ma not change the settings of the set up mode.

    #### TO SET THE ALARM LEVEL

    + 2 sec 2 sec

    Page 17

    Beep high+ Beep low-

    ON/OFF & d

    Switch the pump “on” while holding the “+” ke down. The pump enters the set up mode. First the alarm level can be set: “BEEP HIGH” (+) or “BEEP LOW” (-) Use the “+” or “-” ke to change the setting from “high” to “low” or from “low” to “high”.

    #### TO LOCK THE KEyPAD

    #### +

    Lock Unlk-

    FILL SET d

    Press in the set up mode the “FILL SET” ke to switch to the next setting: “UNLK” the kepad (-) or “LOCK” the kepad (+). In the “LOCK” mode the feeding program of the pump can no longer be changed. Use the “+” or “-” ke to change the setting from “lock” to “unlk” or vice versa.

    #### TO MUTE WHEN DOSE DONE

    #### +

    Beep when done Mute when done-

    FILL SET d

    Press in the set up mode the “FILL SET” ke to switch to the next setting: “BEEP WHEN DONE” (+): the pump will give an alarm when the dose is administered or “MUTE WHEN DONE” (-): the pump will not give an alarm after administering the dose. Use the “+” or “-” ke to change the setting from “beep when done” to “mute when done” or vice versa.

    #### TO SWITCH LIGHT ON

    #### + ON/OFF

    Lite on Lite off-

    FILL SET d d

    2 sec

    Press in the set up mode the “FILL SET” ke to switch to the next setting: “LITE ON” (+) the light of the LCD remains on when the pump is connected to the mains. “LITE OFF” (-) the light of the LCD switches off after a few seconds even if connected to the mains. Use the “+” or “-” ke to change the setting from “lite on” to “lite off” or vice versa.

    #### TO CONNECT WITH NURSE CALL OR PDMS

    #### + ON/OFF

    OUTP PDMS OUTP NRSE-

    FILL SET d d

    2 sec

    Press in the set up mode the “FILL SET” ke to switch to the next setting: “OUTP PDMS” (+) to connect with a Patient Data Monitoring Sstem. “OUTP NRSE” to connect with a NURSE CALL sstem. Use the “+” or “-” ke to change to setting from “OUTP PDMS”, to “OUTP NRSE” or “OUTP OFF” and vice versa.

    Press the “ON/OFF” ke and hold it down for 2 seconds to exit the SET UP mode. The settings are automaticall saved.

    Page 18

    TECHNICAL SPECIFICATIONS

  • • This Flocare® Infinit™ pump in combination with the adapter and the Flocare® Infinit™ Nurse Call are designed to EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for electromagnetic emissions and immunit and are in compliance with Directive 93/42/EEC.
  • 0344

  • • IEC 601: Class II Medical Device Directive: Class IIa
  • • Microprocessor controlled
  • • BF Equipment
  • • Charger: input 100-240V AC / 50-60 Hz / 0.4A Max. output 5 V DC 2.4 A
  • • Batter: internal rechargeable Lithium ion batter, 3.6 V DC 2000 mAh
  • • Batter capacit: 24 hours at 125 ml/h
  • • jet waterproof: IPX 5
  • • IPX 5: Splash water proof: Water ets from an direction shall have no effect on the pump.
  • • Dimensions: 140 x 95 x 35 mm
  • • Weight: approx. 392 g
  • • Accurac flow rate: ±5,0% with appropriate Flocare® Infinit™ pump set
  • • Humidit: Operation mode: 30% to 75% noncondensing Storage: 10% to 95 % noncondensing
  • • Temperature: Operation mode: +5°C to +40°C Storage and transportation: -20°C to +65°C
  • • Atmospheric pressure: Operation mode: 70-106 kPa Storage and transportation: 50-106 kPa
  • • Occlusion detection pressure: Upstream occlusion: -34 kPa (tolerance 21 kPa) Downstream occlusion: 83 kPa (tolerane 21 kPa)
  • • Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activated, varies from 0.5 to 1 ml which relates to an air bubble with an approximate length in the silicone pump segment of 6 13 cm.
  • • The Flocare® Infinit™ can safel be operated on commercial aircraft.
  • • The use of other accessories, adaptors and cables than listed within this manual ma result in increased emissions or decreased immunit of the equipment of the Flocare® Infinit™ pump.
  • • Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the accompaning documents.
  • • Portable and mobile RF communications equipment (cellular telephones) can affect medical electrical equipment. If the Flocare® Infinit™ pump is used adacent to or stacked with other equipment, the pump should be observed to verif normal operation.
  • • In case of pump scrapping, alwas notif our Nutricia sales unit of the pumps serial number. Pump scrapping should alwas occur according to local legislation.
  • APPENDIX A: GUIDANCE AND MANUFACTURER’S DECLARATION ELECTROMAGNETIC COMPATIBILITy

    Guidance and manufacturer’s Declaration – Electromagnetic emissions

    The Flocare® Infinit™ enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Flocare® Infinit™ enteral feeding pump should assure that it is used in such an environment.

    |Emissions test|Compliance|Electromagnetic environment – guidance| |---|---|---| |RF Emissions CISPR 11|Group 1|The Flocare® Infinit™ uses RF energ onl for its internal function. Therefore, its RF emissions are ver low and are not likel to cause an interference in nearb electronic equipment.| |RF Emissions CISPR 11|Class B|The Flocare® Infinit™ is suitable for use in all establishments, including domestic establishments and those directl connected to the public low-voltage power suppl network that supplies buildings used for domestic purposes.| |Harmonic emissions IEC 61000-3-2|Class A|The Flocare® Infinit™ is suitable for use in all establishments, including domestic establishments and those directl connected to the public low-voltage power suppl network that supplies buildings used for domestic purposes.| |Voltage fluctuations/ Flicker emissions IEC 61000-3-3|Complies|The Flocare® Infinit™ is suitable for use in all establishments, including domestic establishments and those directl connected to the public low-voltage power suppl network that supplies buildings used for domestic purposes.|

    Guidance and manufacturer’s Declaration – Electromagnetic immunit

    The Flocare® Infinit™ enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Flocare® Infinit™ enteral feeding pump should assure that it is used in such an environment.

    |Immunit test|IEC 60601 test level|Compliance level|Electromagnetic environment – guidance| |---|---|---|---| |Electrostatic discharge (ESD) IEC 61000-4-2|≤ 6 KV contact ≤ 8 KV air|≤ 6 KV contact ≤ 8 KV air|Floors should be wood, concrete or ceramic tile. If floors are covered with snthetic material, the relative humidit should be at least 30%.|

    Page 19

    |Electric fast transient/burst IEC 61000-4-4|≤ 2 KV for power suppl lines ≤ 1 KV for input/ output lines|≤ 2 KV for power suppl lines ≤ 1 KV for input/ output lines|Mains power qualit should be that of a tpical commercial or hospital environment.| |Surge IEC 61000-4-5|≤ 1 KV differential mode|≤ 1 KV differential mode|Mains power qualit should be that of a tpical commercial or hospital environment.| |Voltage dips, short interruptions and voltage variations on power suppl lines. IEC 61000-4-6|161 Vac for 0,5 s 95 Vac for 100 ms 0 Vac for 10 ms 0 Vac for 5 s|161 Vac for 0,5 s 95 Vac for 100 ms 0 Vac for 10 ms 0 Vac for 5 s|Mains power qualit should be that of a tpical commercial or hospital environment. If the user of the Flocare® Infinit™ requires continued operation during power mains interruptions, it is recommended that the Flocare® Infinit™ enteral feeding pump be powered from an uninterruptible power suppl of batter.|

    ##### Recommended separation distances between portable and mobile RF communications euipment and the Flocare

    ##### Infinit

    ##### enteral feeding pump

    ®

    The Flocare® Infinit™ enteral feeding pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Flocare® Infinit™ can help prevent electromagnetic interference b maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Flocare® Infinit™ as recommended below, according to the maximum output power of the communications equipment.

    |Rated Maximum output power of transmitter (W)|Separation distance according to frequenc of transmitter (m)|Separation distance according to frequenc of transmitter (m)|Separation distance according to frequenc of transmitter (m)| |---|---|---|---| |Rated Maximum output power of transmitter (W)|150 KHz to 80 MHz d 0,12CDP|80 MHz to 800 MHz d 0,12CDP|800 MHz to 2,5 GHz d 0,23CDP| |0,01|0,01|0,01|0,02| |0,1|0,04|0,04|0,07| |1|0,12|0,12|0,23| |10|0,37|0,37|0,74| |100|1,17|1,17|2,33|

    For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequenc of the transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to the transmitter manufacturer.

  • NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequenc range applies.
  • NOTE 2 These guidelines ma not appl in all situations. Electromagnetic propagation is affected b absorption and reflection from structures, obects and people.
  • ##### Guidance and manufacturer’s Declaration – Electromagnetic immunit

    enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Flocare

    The Flocare

    Infinit

    ®

    Infinit

    enteral feeding pump should assure that it is used in such an environment.

    ®

    |Immunit test|IEC 60601 test level|Compliance level|Electromagnetic environment – guidance| |---|---|---|---| |Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3

    |3 Vrms 150 KHz to 80 MHz 3 V/m 80 Mhz to 2,5 GHz

    |3 Vrms 150 KHz to 80 MHz 30 V/m 80 Mhz to 2,5 GHz

    |Portable and mobile RF communications equipment should be used no closer to an part of the Flocare® Infinit™, including cables, than the recommended separation distance calculated from the equation applicable toe the frequenc of the transmitter: Recommended separation distance

    Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined b an electromagnetic site surve,a should be less than the compliance level in each frequenc range.b Interference ma occur in the vicinit of equipment marked with the following smbol:

    d = 1,17CDP d = 0,12CDP d = 0,23CDP

    80 MHz to 800 MHz 800 MHz to 2,5 GHz|

  • NOTE 1 At 80 Mhz and 800 MHz, the higher frequenc range applies.
  • NOTE 2 These guidelines ma not appl in all situations. Electromagnetic propagation is affected b absorption and reflection from structures, obects and people.
  • a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoreticall with accurac. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site surve should be considered. If the measurement field strength in the location in which the Flocare® Infinit™ is used exceeds the applicable RF compliance above, the Flocare® Infinit™ should be observed to verif normal operation. If abnormal performance is observed, additional measures ma be necessar, such as reorienting or relocating the Flocare® Infinit™.
  • b. Over the frequenc range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
  • Similar Unknown